Model Number 25-40-45 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/15/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).The device has been returned to livanova (b)(4) for investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that an s5 gas blender system displayed an error message during maintenance.There was no patient involvement.
|
|
Manufacturer Narrative
|
The s5 gas blender system was returned to manufacturer for further investigation.During the challenging testing the reported failure could not be confirmed or reproduced.The device worked according to the specification.A new calibration and subsequent functional verification testing was completed without issues.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
|
|
Event Description
|
See initial report.
|
|