| Model Number |
CRPLUS |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer received a replacement device.Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that during a patient event, their device would not detect that a patient was connected.The device gave a "connect electrodes" message even though the patient was connected to the device.Medical personnel removed the electrodes from the device and placed them into a backup device when ems arrived on scene.There was no adverse effects to the (b)(6) year old patient as a result of the reported issue as the backup device was used to continue patient care.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and verified the reported issue.The therapy harness connection to the upper case was not connected (manufacturing error), and a patient load could not be recognized.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that during a patient event, their device would not detect that a patient was connected.The device gave a "connect electrodes" message even though the patient was connected to the device.Medical personnel removed the electrodes from the device and placed them into a backup device when ems arrived on scene.There was no adverse effects to the (b)(6) patient as a result of the reported issue as the backup device was used to continue patient care.
|
| |
|