| Model Number |
MMT-723LNAP |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed button error and battery out limit.Customer's blood glucose level was not reported.The buttons were unresponsive.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Pump passed displacement test, self test, unexpected restart error test and all operating currents tested within the specifications range.Pump was monitored and tested with no unexpected battery out limit alarm noted.Pump received with corroded battery tube noted.
|
| |
|