Catalog Number CRUOS106
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records has reviewed.The device was returned to the manufacturer for evaluation.The investigation is currently under way.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during the use of a recanalization catheter in the posterior tibial, the distal tip of the recanalization catheter allegedly separated from the catheter but remained attached to the core wire.Reportedly, another recanalization catheter was used to complete the procedure.There was no reported patient injury.
 
Event or Problem Description
It was reported that during the use of a recanalization catheter in the posterior tibial, the distal tip of the recanalization catheter allegedly separated from the catheter but remained attached to the core wire.Reportedly, another recanalization catheter was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.One crosser catheter was returned for evaluation.The outer catheter of the returned device was noted to be separated from the distal tip of the device.The marker band was noted to be in place on the outer catheter.Therefore, based on the returned device, the investigation is confirmed for material separation.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER CTO RECANALIZATION CATHETER
Common Device Name
RECANALIZATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7589591
Report Number2020394-2018-00842
Device Sequence Number2282032
Product Code PDU
UDI-Device Identifier10801741125611
UDI-Public(01)10801741125611
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K161208
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2020
Device Catalogue NumberCRUOS106
Device Lot NumberGFCQ3699
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Initial Date Received by Manufacturer 05/16/2018
Supplement Date Received by Manufacturer08/01/2018
Initial Report FDA Received Date06/11/2018
Supplement Report FDA Received Date08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
Patient Weight93
-
-