| Catalog Number |
3L92516 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Pain (1994); Tissue Damage (2104); Test Result (2695); Not Applicable (3189)
|
| Date of Event |
03/22/2014
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary: the asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.The investigation regarding the root cause(s) and/or corrective actions was conducted under mdd capa (b)(4).Ongoing post market surveillance is conducted per our procedures for this product.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Update (b)(6) 2018: asr medical records received. after review of medical records, the patient was revised due to aseptic loosening of the left hip joint endoprosthesis acetabulum, metallosis, and femur fracture.Revision notes reported the head was broken away, stable stem, and loose acetabulum removed.Lab result shows metal ion levels above 7 ppb. doi: (b)(6) 2009; dor: (b)(6) 2014; left hip. .
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary: the asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.The investigation regarding the root cause(s) and/or corrective actions was conducted under mdd capa-(b)(4).Ongoing post market surveillance is conducted per our procedures for this product.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|