Model Number TRUETRACK
Medical Device Problem Code Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Test strips not returned for evaluation.Returned meter evaluated with defect found.On the investigation on 5/30/2018 resulted in defect found.Meter displays partial characters.Based on the result the event was determined to be reportable.Final root cause : meter displays partial characters due to user mechanical stress.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Message left to contact customer care department.
 
Event or Problem Description
Consumer complaint of defective lcd.Customer feels well and observed no symptoms.Medical intervention is not reported at the time of the call on (b)(6) 2018 as a result of the meter's readings.Storage of product location is undisclosed.Test strip lot manufacturer's expiration date is 08/13/2019 and open vial date is undisclosed.Customer states that the middle digit is illegible, it is not clear and cannot read it.Caller states the first and third digit are legible.No medication was administered due to the glucose reading.No adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7591215
Report Number1000113657-2018-00641
Device Sequence Number1698851
Product Code NBW
UDI-Device Identifier00021292001018
UDI-Public(01)00021292001018
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRU5088
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2018
Initial Report FDA Received Date06/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-