Catalog Number 305784
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the after drawing up medication using the bd eclipse¿ needles - bd luer-lok¿ syringe, a leakage occurred at the threaded junction point.Found during use.No reports of serious injury or medical intervention noted.
 
Additional Manufacturer Narrative
3 pcs representative samples were received in sealed package.All the samples were subjected to leak test and passed the test.A device history record review showed no non-conformances associated with this issue during the production of this batch.3 pcs representative samples were returned and subjected to leak test.All samples passed the test.Hence, root cause could not be determined.Continue to monitor the nonconformance trend.The representative samples passed the leak test.Continue to monitor the nonconformance trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ECLIPSE¿ NEEDLES - BD LUER-LOK¿ SYRINGE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
MDR Report Key7591338
Report Number8041187-2018-00188
Device Sequence Number16618859
Product Code FMI
UDI-Device Identifier00382903057849
UDI-Public00382903057849
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K161170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/28/2023
Device Catalogue Number305784
Device Lot Number8052065
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2018
Initial Date Received by Manufacturer 05/22/2018
Supplement Date Received by Manufacturer05/22/2018
Initial Report FDA Received Date06/12/2018
Supplement Report FDA Received Date06/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-