• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM Back to Search Results
Catalog Number 1012583-16
Health Effect - Clinical Code Death (1802)
Date of Event 05/17/2018
Type of Reportable Event Death
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported failure to seal the perforation.A conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined.In addition, the reported patient effect of death is listed in the graftmaster rapid exchange (rx) coronary stent graft system, domestic instructions for use (ifu), as a known patient effect of coronary stenting procedures.It was reported that the graftmaster was deployed with a maximum pressure of 14 atmospheres (atm).It should be noted in the graftmaster rapid exchange (rx), coronary stent graft system, domestic, instructions for use (ifu) specifies: deploy the stent graft slowly by pressurizing the delivery system in 2-atm increments, every 5 seconds, until the stent graft is completely expanded.Fully expand the stent graft by inflating to nominal pressure at a minimum.The ifu, table 10 specifies the nominal pressure is 15 atmospheres (atm).It is unknown if the ifu deviation contributed to the reported event.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.The other graftmaster devices referenced in describe event or problem is being filed under a separate medwatch report.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the proximal mid left anterior descending (lad) artery.The patient presented with heart failure and non-st-elevation myocardial infarction (nstemi), and was deemed not to be a surgical candidate.The perforation could not be sealed after five graftmaster stents were deployed at maximum pressure (14 atmospheres).A 4.5 x 16 mm and a 3.5 x 16 mm graftmaster stent were implanted to seal the perforation, but the stents could not seal the large perforation, and therefore another 3.5 x 16 mm and a 2.5 x 16 mm graftmaster stents were implanted inside the first two stents to help seal the perforation.A decision was made to implant a final stent, a 2.5 x 16 graftmaster stent to ensure the target lesion was fully covered.However, the patient's vessels were so fragile that the proximal and mid lad split open after rotational atherectomy.The physician reported that the perforation was too large and the patient expired as a result of the rotational atherectomy, and not the graftermaster stents.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Common Device Name
CORONARY STENT GRAFT
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7591387
Report Number2024168-2018-04385
Device Sequence Number14008725
Product Code MAF
UDI-Device Identifier08717648176470
UDI-Public08717648176470
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
H000001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2018
Device Catalogue Number1012583-16
Device Lot Number6090141
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2018
Initial Report FDA Received Date06/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age81 YR
Patient Weight102
-
-