COVIDIEN LP LLC NORTH HAVEN IDRIVE ULTRA; STAPLE, IMPLANTABLE Back to Search Results
Model Number IDRVULTRA2
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a robotic laparoscopic lower anterior resection, on the rectum tissue, the stapler was not able to fire.The surgeon was able to press the green button and squeeze the handle but the jaws of the device lock on tissue and was removed by opening it.The surgical time was extended by thirty minutes or more and the surgeon had to switch to manual gun in order to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.No visual abnormalities were noted.The electrically erasable programmable read-only memory (eeprom) was uploaded and indicated 11 autoclave cycles for the handle.A pmv representative adapter and single use loading unit (sulu) were connected to the subject handle and applied to test media.The device responded correctly and the reload was able to be fired.The reload cleanly applied staples and transected the test media.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDRIVE ULTRA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7591416
Report Number1219930-2018-03039
Device Sequence Number18900772
Product Code GDW
UDI-Device Identifier10884521510579
UDI-Public10884521510579
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K121510
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIDRVULTRA2
Device Catalogue NumberIDRVULTRA2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2018
Initial Date Received by Manufacturer 05/26/2018
Supplement Date Received by Manufacturer08/16/2018
Initial Report FDA Received Date06/12/2018
Supplement Report FDA Received Date09/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-