| Catalog Number |
PTA3-14-150-2-12 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/27/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
|
| |
|
Event or Problem Description
|
|
The complainant reported a cross-over pta (percutaneous transluminal angioplasty) in the crural vessels.Another manufacturer's wire guide was stuck inside the advance micro 14 ultra low-profile pta balloon catheter during recanalization of the crural vessels.It was further noted that the wire guide remained inside the balloon catheter, and a portion of the catheter felt stretched.There were no reported serious injuries to the patient as a result of the product problem.
|
| |
|
Search Alerts/Recalls
|