• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ANGIODYNAMICS / BIOFLO; PERIPHERALLY INSERTED CENTRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ANGIODYNAMICS / BIOFLO; PERIPHERALLY INSERTED CENTRAL CATHETER Back to Search Results
Catalog Number H965458950
Medical Device Problem Code Physical Property Issue (3008)
Health Effect - Clinical Codes Fever (1858); Unspecified Infection (1930)
Date of Event 05/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation into this event is ongoing, both at angiodynamics and at (b)(4).Upon completion of the investigation a supplemental medwatch will be submitted.(b)(4).
 
Event or Problem Description
Angiodynamics was notified by (b)(4), the supplier of introducer sheaths, that they have initiated a recall (1649395-05-02-2018-001-r) on specific device lots due to "unsafe levels of bacterial endotoxins (pyrogens) that were introduced during a manufacturing step." (b)(4) has informed angiodynamics that use of the introducers may expose patients/end users to "bacterial endotoxins, also called pyrogenic bacteria that can activate the inflammatory process and produce fever, chills, and hypotension in a patient." as a result of the (b)(4) recall, angiodynamics is conducting a voluntary recall (removal) of the affected angiodynamics products.The recall notification letters were delivered to affected consignees on friday may 11, 2018.After receiving notification, the end user hospital has reported: "patient exhibited potential infectious symptoms the next day.No treatment required.(b)(6) - increased temp one time to 38.6 (101.5) despite antibiotics with c/o chills, sweats, generalized aches and ha patient had now recalled picc line device placed (b)(6) 2018.
 
Additional Manufacturer Narrative
In lieu of a reported lot number, a ship history report (shr) was generated for item number (h965458950) in order to determine the last three lots shipped to the reporting hospital in the six months prior to the procedure date.The three lots obtained through the shr were (5295311, 5304321 and 5312589).The device history records for the lots obtained through the ship history report (packaging lots) were reviewed.In addition, the corresponding lots for purchased component item number 06509208 (angiodynamics lot 639464 {supplier lot number 17262440}, 640497 {supplier lot number 17356436} and 640330 {supplier lot number 17321977} packaged in it.A review of the device history records was performed for the indicated packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The recent angiodynamics complaint report was reviewed for the bioflo picc product family and the failure mode "patient injury - infection." no adverse trends were indicated.The end user hospital's reported complaint description cannot be confirmed given the nature of the adverse event.The customer was reviewing patient's affected by the galt sheath recall and observed infection post placement.Angiodynamics received bio-testing results of galt manufacturing process performed by nelson labs.Only one organism present has potential to produce endotoxin reaction.The relationship between the galt sheath and the patient infection cannot be definitively determined by angiodynamics.(b)(6) has initiated a capa to address the issue regarding recall # z-2196-2018.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS / BIOFLO
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key7591592
Report Number1317056-2018-00110
Device Sequence Number14002083
Product Code LJS
UDI-Device IdentifierH965458950
UDI-PublicH965458950
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K133264
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Risk Manager
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Catalogue NumberH965458950
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2018
Supplement Date Received by Manufacturer06/08/2018
Initial Report FDA Received Date06/12/2018
Supplement Report FDA Received Date09/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2196-2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
Patient Weight82
-
-