Angiodynamics was notified by (b)(4), the supplier of introducer sheaths, that they have initiated a recall (1649395-05-02-2018-001-r) on specific device lots due to "unsafe levels of bacterial endotoxins (pyrogens) that were introduced during a manufacturing step." (b)(4) has informed angiodynamics that use of the introducers may expose patients/end users to "bacterial endotoxins, also called pyrogenic bacteria that can activate the inflammatory process and produce fever, chills, and hypotension in a patient." as a result of the (b)(4) recall, angiodynamics is conducting a voluntary recall (removal) of the affected angiodynamics products.The recall notification letters were delivered to affected consignees on friday may 11, 2018.After receiving notification, the end user hospital has reported: picc associated with (b)(4) sheath recall (for bacterial endotoxins) was placed in a patient on (b)(6) 2018 in the medsurg unit.At 2300 hours, the patient was transferred to icu with a fever of 103, hypoxia, and hypotension.They drew cultures from the picc and a separate peripheral site.Eight hours post results were negative.(b)(6) - patient had many complications from rectal cancer and now needed tpn.Picc placed approximately 1100.At 11pm, per notes, patient developed respiratory distress necessitating transfer to icu and intubation.Patient also experienced fever to 103.0 and hypotension necessitating dopamine which was later switched to norepinephrine.Patient was bronched and cultures drawn from picc as well as peripheral stick.Bal performed with bronch that so far shows bacteriology preliminary report : may 15, 2018 14:42 tech code: 244.Gram stain: 4+ wbc's, 2+ epithelial cells, 1+ phagocytized gram positive cocci in clusters, 4+ diphtheroid-like gram positive rods.Cxr has been suggestive of pneumonia.Patient now currently stable and was extubated today.Pressors have been weaned but still necessary.(b)(6): blood cultures drawn from picc as well as peripheral stick have remained negative through final read.Patient now on minimal o2 requirements.
|