• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® CLEAN-CATH® CATHETER; VINYL FEMALE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® CLEAN-CATH® CATHETER; VINYL FEMALE CATHETER Back to Search Results
Model Number 420614
Medical Device Problem Code Misassembled (1398)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that the catheter was placed backwards inside of the packaging.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing related.Received 6 unopened clean-caths for evaluation.The visual evaluation of the returned samples noted six unopened (within original packaging present), unused vinyl catheters present.The drainage eye end of the catheters was facing the chevron seal.This configuration was out of specification, which states the funnel of the catheter must be in the chevron seal of the package.No functional requirements were specified.All six catheters were removed from their sealed packaging.Inner diameter, outer diameter, french size, and catheter length were measured for all six catheters and all catheter dimensions measured were found to be within specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿bard clean-cath sterile catheter for intermittent catheterization.Visually inspect the product for any imperfections or surface deterioration prior to use.For urological use only.To use: insert rounded end of catheter.Caution: federal (usa) law restricts this device to sale by or on the order of a physician.Contains or presence of phthalates: di(2-ethylhexyl)phthalate (dehp) is a plasticizer used in some polyvinyl chloride medical devices.Dehp has been shown to produce a range of adverse effects in experimental animals, notably liver toxicity and testicular atrophy.Although the toxic and carcinogenic effects of dehp have been well established in experimental animals, the ability of this compound to produce adverse effects in humans is controversial.There is no evidence that neonates, infants, pregnant and breast feeding women exposed to dehp experience any related adverse effects.However, a lack of evidence of causation between dehp-pvc and any disease or adverse effect does not mean that there are no risks.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.This is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of device, which may lead to device failure, and/or lead to injury, illness or death of the patient".
 
Event or Problem Description
It was reported that the catheter was placed backwards inside of the packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® CLEAN-CATH® CATHETER
Common Device Name
VINYL FEMALE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7591884
Report Number1018233-2018-02201
Device Sequence Number10833899
Product Code KOD
UDI-Device Identifier00801741038921
UDI-Public(01)00801741038921
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K802665
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/31/2022
Device Model Number420614
Device Catalogue Number420614
Device Lot NumberNGBZ1467
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2018
Initial Date Received by Manufacturer 05/23/2018
Supplement Date Received by Manufacturer09/10/2018
Initial Report FDA Received Date06/12/2018
Supplement Report FDA Received Date09/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-