BARD PERIPHERAL VASCULAR, INC. FLAIR ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FAF09050
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device has returned.The investigation is currently under way.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a stent graft placement procedure, the tip of the catheter allegedly broke off while the stent was loading over the wire.Reportedly, another endovascular stent was used to complete the procedure.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a stent graft placement procedure, the tip of the catheter allegedly broke off while the stent was loading over the wire.Reportedly, another endovascular stent was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.During visual inspection the atraumatic tip was detached from the delivery system with a portion of the guidewire lumen and marker distal band attached.Therefore the investigation is confirmed for detachment and unconfirmed for the reported break as the distal segment was completely separated from the rest of the delivery system.The definitive root cause for the identified issues could not be determined based upon available information.It is unknown if procedural issues contributed to the event.Labeling review:the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Eval code & desc - conclusion1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLAIR ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7592046
Report Number2020394-2018-00846
Device Sequence Number19459727
Product Code PFV
UDI-Device Identifier04049519008875
UDI-Public(01)04049519008875
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P060002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/13/2020
Device Model NumberFAF09050
Device Catalogue NumberFAF09050
Device Lot NumberANBN1429
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2018
Initial Date Received by Manufacturer 05/16/2018
Supplement Date Received by Manufacturer09/17/2018
Initial Report FDA Received Date06/12/2018
Supplement Report FDA Received Date09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age48 YR
Patient Weight104
-
-