• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523NAB
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 01/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The insulin pump was received with a blank display due to moisture damage at the electronic assembly.Moisture damage was also identified on the motor.The blank display prevented functional testing, which includes: self test, rewind test, displacement test, and verification of a low battery alarm.The following physical damage was noted: minor scratched lcd window, cracked case at the display window corners, cracked lcd window, cracked belt clip slot, cracked battery tube threads, stained end cap sticker, stained address/serial number label and broken reservoir tube lip and missing reservoir tube o ring.A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.
 
Event or Problem Description
It was reported via phone call that their insulin pump kept turning off and had to replace the battery frequently.It was also stated there was a broken piece at the reservoir compartment.The customer¿s blood glucose level was 411 mg/dl at the time of the incident, came down to 312 mg/dl and 175 mg/dl after the correction.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7592609
Report Number3004209178-2018-80123
Device Sequence Number19276390
Product Code OYC
UDI-Device Identifier00643169202306
UDI-Public(01)00643169202306
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAB
Device Catalogue NumberMMT-523NAB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2018
Initial Report FDA Received Date06/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-