MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number M004PM4790K20
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported foreign material was found inside the sterile package.During preparation of the intellatip mifi¿ xp ablation catheter, an object which appeared to be ¿trash¿ was found inside the sterile inner package.The procedure was completed with a different catheter.There were no patient complications.
 
Event or Problem Description
It was reported foreign material was found inside the sterile package.During preparation of the intellatip mifi¿ xp ablation catheter, an object which appeared to be ¿trash¿ was found inside the sterile inner package.The procedure was completed with a different catheter.There were no patient complications.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the device was returned inside of the original sealed pouch, with the original opened box.Visual inspection confirmed foreign material, measuring greater than 5 square millimeters, moving freely within the sealed pouch.The investigation conclusion is manufacturing execution error as the manufacturing process was not executed as validated/as designed.Updated: device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
MDR Report Key7592829
Report Number2134265-2018-05147
Device Sequence Number10831841
Product Code OAD
UDI-Device Identifier08714729848059
UDI-Public08714729848059
Combination Product (Y/N)N
PMA/510(K) Number
P020025
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model NumberM004PM4790K20
Device Catalogue NumberPM4790K2
Device Lot Number0021456436
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2018
Initial Date Received by Manufacturer 05/21/2018
Supplement Date Received by Manufacturer07/03/2018
Initial Report FDA Received Date06/12/2018
Supplement Report FDA Received Date07/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-