| Model Number |
M004PM4790K20 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported foreign material was found inside the sterile package.During preparation of the intellatip mifi¿ xp ablation catheter, an object which appeared to be ¿trash¿ was found inside the sterile inner package.The procedure was completed with a different catheter.There were no patient complications.
|
| |
|
Event or Problem Description
|
|
It was reported foreign material was found inside the sterile package.During preparation of the intellatip mifi¿ xp ablation catheter, an object which appeared to be ¿trash¿ was found inside the sterile inner package.The procedure was completed with a different catheter.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned inside of the original sealed pouch, with the original opened box.Visual inspection confirmed foreign material, measuring greater than 5 square millimeters, moving freely within the sealed pouch.The investigation conclusion is manufacturing execution error as the manufacturing process was not executed as validated/as designed.Updated: device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes.(b)(4).
|
| |
|