| Model Number |
Z-800 |
| Medical Device Problem Codes |
Inaccurate Flow Rate (1249); Device Operates Differently Than Expected (2913); Electrical Power Problem (2925)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump was found to power on as intended, and the issue of fluctuating flow rate was not witnessed.Pump was found to have a flow rate accuracy of -4.76% which is within the +/-5% flow rate specification.The reported issues were not confirmed.
|
| |
|
Event or Problem Description
|
|
The customer reported flow rate issue and power issue with the device."the pump just flashed and changed the rate in 2 seconds, and then completely wiped it to zero." "it beeps when off.Turns itself off then on again and changes the rates on its own." no patient was involved in this case.
|
| |
|