BARD PERIPHERAL VASCULAR, INC. DORADO PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number DR80610
Medical Device Problem Codes Break (1069); Inflation Problem (1310)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or no medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that when inflating a pta balloon to 22 atm, the balloon allegedly was unable to inflate to the desired pressure, not exceeding 20 atm.It was further alleged that an attempt to inflate the balloon to 12 atm was successful.Reportedly, upon removal, damage was found.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Investigation summary: the device was returned for evaluation.The device was inflated with water, and water was seen exiting the device at the proximal balloon butt joint.The device was unable to be fully inflated.Upon closer examination, a partial break at the butt joint was identified.Therefore, the investigation is confirmed for a break at the butt joint as well as confirmed for the reported inflation issue.The partial break at the glue butt joint resulted in the inflation issue.However, the definitive root cause for the partial break could not be determined based upon available information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that when inflating a pta balloon to 22 atm, the balloon allegedly was unable to inflate to the desired pressure, not exceeding 20 atm.It was further alleged that an attempt to inflate the balloon to 12 atm was successful.Reportedly, upon removal, damage was found.There was no reported patient injury.
 
Event or Problem Description
It was reported that when inflating a pta balloon to 22 atm, the balloon allegedly was unable to inflate to the desired pressure, not exceeding 20 atm.It was further alleged that an attempt to inflate the balloon to 12 atm was successful.Reportedly, upon removal, damage was found.There was no reported patient injury.
 
Additional Manufacturer Narrative
No medical records or no medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The previously submitted emdr inadvertently left blank.The date on the emdr should have been reported as 05/14/2018.Section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7593090
Report Number2020394-2018-00852
Device Sequence Number18884454
Product Code DQY
UDI-Device Identifier00801741058837
UDI-Public(01)00801741058837
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDR80610
Device Lot Number93YB0048
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer07/03/2018
08/01/2018
Initial Report FDA Received Date06/12/2018
Supplement Report FDA Received Date07/24/2018
08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
Patient Weight65
-
-