This information is unknown.Date of pars fracture is unknown.Date of implantation is unknown date in 2013.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter email address is unknown.The investigation could not be completed; no conclusion could be drawn, as no product was received.A device history records review has been requested.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was report that the patient was originally implanted sometime in 2013 with a prodisc-l (pdl) total disc replacement at spine disc levels l4-l5.Patient was implanted with one (1) pdl construct.The construct included a superior end plate large, polyethylene inlay large and inferior end plate large.On an unknown date post-operative, the patient presented with a pars fracture which caused the patient consistent pain.The pars fracture was located behind the pdl l4-l5 implant.It is unknown what caused the pars fracture.On (b)(6) 2018, the surgeon removed all implants and revised the patient to a competitor¿s anterior fusion with cage.Inter-operative, due to patient's scar tissue, the pdl superior end plate proved difficult to remove.Revision surgery was completed successfully.No fragments were generated during implant removal.The patient is reported in stable condition.This report is for a superior end plate large 6°-sterile.This is report 1 of 3 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: device history record (dhr) review: part number pdl-l-sp06s; lot number: 6816821; date of manufacture: february 08, 2012; place of manufacture: brandywine plant.Part expiration date: 12/2016.No nonconformance records (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Investigation summary: because there was no product problem reported and the reported damage is consistent with implant, use and explant of the device the ¿adverse event (no reported product problem) flow was chosen.The prodisc-l total disc replacement is indicated for spinal arthroplasty in skeletally mature patients with degenerative disc disease (ddd) at one level from l3 to s1.The complaint description stated that the implants were removed with the goal of replacement with a fusion device.The exact reason for removal is unknown.No product problems were reported.One prodisc-l superior endplate (item pdl-l-sp06s lot 6816821), polyethylene inlay (item pdl-l-pt14s lot 68396**), and inferior endplate (pdl-l-ip00s lot 6868300) were returned for analysis.The drawings for the superior endplate, inferior endplate and polyethylene inlay were reviewed.Visual examination did not reveal any design related issues.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.As received condition- superior end plate, pdl-l-sp06s, lot #6816821 was received with evidence of iatrogenic damage that includes on top face a dimpled imprint, also, there is evidence of impingement indicated by micro-lines on radius posterior side off the spherical diameter, the polished spherical diameter has a few minuscule scratches, and the instrument face with two holes has dent on both sides of holes.The back cpti plasma spray surface has remnants of bony tissue on both sides of the keel, the keel and barbs and spikes.The anterior edge of this surface has nicks and scratches with exposed metal.There are two areas parallel to the keel where the cpti plasma spray surface has been visually flattened/shiny.All described nonconformities are post manufacturing.Summary- the superior end plate was reviewed for this complaint.The event description states ¿it is unknown what caused the pars fracture.¿ so, from the manufacturing perspective there are no complaints stated against this product; therefore, there will be no further investigation.Complaint is not confirmed from a manufacturing perspective.Review of exponent report- the devices were reviewed in exponent (18050901) report.This complaint ¿as received condition¿ was compared to the third-party evaluation by ¿exponent¿ and found similar conditions.The following were noted by exponent.The implants were returned with damage that is consistent with removal of the implants.The backside surface of the superior and inferior endplates had remnants of bone.Polyethylene inlay had evidence of impingement.Evidence of worn machining marks on the backside surface of the polyethylene inlay.The articulating surface of the polyethylene inlay has areas of discoloration with burnishing and multidirectional micro scratches.The snap lock of the polyethylene inlay appeared rounded when compared to a never implanted control.Specification- there are no specifications to address, there are no complaints stated against this product.Because the reported complaint does not indicate a product problem or surgical technique issue and no product problems were reported during investigation, a risk document review will not be conducted.No design or manufacturing issues were identified during investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|