| Model Number |
NS2000 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Uterine Perforation (2121)
|
| Date of Event |
05/10/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not being returned therefore, a failure analysis of the complaint device cannot be completed.Standard dhr review|a device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an uncomplicated novasure endometrial ablation was performed by a physician on (b)(6) 2018.The next day the patient was "admitted and underwent surgery including a bowel resection".No additional information was received.
|
| |
|
Event or Problem Description
|
|
"the patient called 24 hours later with rigors that spontaneously resolved.At 36 hours post-procedure, she went to the er with fever.A ct-scan showed abnormal changes in the myometrium along with fluid and free air in the pelvis.She was taken to the or for a laparotomy.Upon entry into the abdomen, bowel contents were noted.A portion of bowel was adherent to the posterior wall of the uterus.The bowel was dissected off of the uterine wall; blanching of the uterine serosa was noted along the posterior fundus consistent with thermal injury.A thermal injury with perforation was identified on the small bowel at the distal ileum.A resection with primary anastomosis was performed.The patient was discharged home in stable condition after one week of hospitalization".
|
| |
|
Additional Manufacturer Narrative
|
|
The radio frequency controller (rfc) was returned on 06/07/2018: rf controller s/n (b)(4) passed all functional testing.This observation will be monitored and trended.Standard dhr review|a device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.
|
| |
|