• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LONG NAIL KIT R2.0, TI, RIGHT GAMMA3® Ø11X320MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LONG NAIL KIT R2.0, TI, RIGHT GAMMA3® Ø11X320MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 32250320S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/04/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device disposition is unknown.
 
Event or Problem Description
It was reported by the end user that 3 weeks after the medical procedure, while the patient was walking, she heard a 'crack' noise from her hip.At the hospital they found that the nail was broken.The nail was replaced on another surgery.
 
Event or Problem Description
It was reported by the end user that 3 weeks after the medical procedure, while the patient was walking, she heard a 'crack' noise from her hip.At the hospital they found that the nail was broken.The nail was replaced on another surgery.
 
Additional Manufacturer Narrative
Concomitant medical products the reported event that long nail kit r2.0, ti, right gamma3® ø11x320mm x 125° was alleged of k-170 (implant breakage - nail) could be confirmed, since the device returned for evaluation matches the alleged failure mode.Visual inspection revealed that the evaluation revealed that the nail broke in a fatigue fracture after a period of more than 8 months of implantation.According to the inspection significant material damage had weekend the web caused by misaligned drilling with the step drill; which presents an unintentional use error.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Thus, based on the investigation the root cause can most likely be attributed to ¿user related¿, as the nail was considerably pre-damaged intra-operatively most likely due to misaligned drilling with the step drill; which presents an unintentional use error.If any further information is provided, the investigation report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R2.0, TI, RIGHT GAMMA3® Ø11X320MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
MDR Report Key7593473
Report Number0009610622-2018-00239
Device Sequence Number1317292
Product Code HSB
UDI-Device Identifier07613153312734
UDI-Public07613153312734
Combination Product (Y/N)N
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/30/2022
Device Catalogue Number32250320S
Device Lot NumberK05B834
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2018
Initial Date Received by Manufacturer 05/21/2018
Supplement Date Received by Manufacturer10/11/2018
Initial Report FDA Received Date06/12/2018
Supplement Report FDA Received Date11/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-