| Model Number |
4FC12 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
No Code Available (3191); Pericardial Effusion (3271)
|
| Date of Event |
05/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cryo ablation procedure, an anterior moderate pericardial effusion was noticed.The physician discontinued heparin, removed balloon and sheath into the right heart, and did immediate pericardiocentesis.Intracardiac echo showed removal of 550 ml of fluid.Protamine was given, ablation was not resumed.The adverse event resolved and the patient was discharged from the hospital with no signs of residual effusion.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information reported that the hospital stay was not extended.An echocardiogram the following day showed no effusion and the patient had a follow up echo a week later which showed no effusion.The case was aborted and the patient was under general anesthesia.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: this event is a duplicate of fda report number 3002648230-2018-00376.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|