| Model Number |
ZCT300 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
| Date of Event |
04/09/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If explanted, give date: not applicable, as the lens remains implanted.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zct300 26.0 diopter lens was implanted in the patient's left eye (os) on (b)(6) 2018 at an axis of 002 degrees.Post-operatively on (b)(6) 2018, at the patient's follow up visit, the patient had blurred vision secondary to a shift in the cornea which created a shift in the axis of cylinder causing the lens axis to change to 010 degrees.Therefore, a lens repositioning procedure was performed on (b)(6) 2018.Reportedly, there was no patient complications and/or related injury.Outcomes did not significantly interfere with activities of daily life.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site as to date it remains implanted; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received which reported the subject came in for a 6-month study visit on (b)(6) 2018 and the intraocular lens (iol) axis was measured as follows: os: 165 degrees (9 degrees from iol repositioned location of 156 degrees).Iol reposition occurred on (b)(6) 2018.Currently there is no planned intervention for the new rotation.
|
| |
|