| Model Number |
CQ2015A |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: collamer ultraviolet-absorbing posterior chamber three piece foldable intraocular lens the lens was returned in liquid, in lens case/vial.Visual inspection found the lens optic torn and one haptic bent.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated there was a loading issue with a cq2015a collamer three piece lens, +20.0 diopter.There was patient contact but no injury noted.The backup lens was implanted.The reporter indicated the surgeon used a competitor's injection system.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Work order search: no similar complaint type events were reported for units within the same lot.Corrected data: claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: device evaluation: visual inspection found the lens optic and haptic torn; haptic bent.Claim# (b)(4).
|
| |
|