| Catalog Number |
366664 |
| Medical Device Problem Code |
Incorrect Or Inadequate Test Results (2456)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/09/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode with the incident lot was not observed.A review of the manufacturing records was completed for the incident lot and no issues were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer® lithium heparin (lh) 37 usp units blood collection tubes had "erroneous results." no serious injury or medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: bd is issuing a voluntary medical device correction for bd vacutainer® plastic lithium heparin tubes.False elevation of carboxyhemoglobin (cohb) results have been reported when analyzing venous blood samples collected with bd vacutainer® plastic lithium heparin tubes with the il gem 4000 instrument.Please reference bd recall #: pas-18-1175-fa.
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer® lithium heparinn (lh) 37 usp units blood collection tubes had "erroneous results." no serious injury or medical intervention.
|
| |
|