| Catalog Number |
M0035423100 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not available to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that when the coil was advanced in the microcatheter, the coil detached prematurely inside the microcatheter.The coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record review confirms that the device met all material, assembly and performance specifications.Visual inspection of the device revealed that the delivery wire was broken/fractured and the main coil was kinked.Functional testing could not be performed due to the damaged condition of the returned coil.Device analysis revealed that the main coil had not detached as originally reported, but the coil delivery wire was broken.Based on the information currently available, it is probable that some procedural factors encountered during the procedure limited the performance of the coil contributing to the reported and observed damages.Therefore, an assignable cause of component failure has been assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that when the coil was advanced in the microcatheter, the coil detached prematurely inside the microcatheter.The coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
|
| |
|