Catalog Number 306545
Health Effect - Clinical Code Bacterial Infection (1735)
Date of Event 04/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.Investigation: no samples were provided for evaluation.A device history review could not be completed as no batch number was provided.Controls in place at the manufacturing site: bioburden tested on a weekly basis.An overkill sterilization process is used.Environmental testing within the filling area is done on a biweekly basis.The sterilization process is challenged and re-qualified annually.The solution is filtered twice before it is filled in syringes (note: once when transferring to the hold tank and again at the fill machine).Endotoxin testing is performed on each batch.Fill room operators are trained and qualified on gowning for the fill room environment on an annual basis.Weekly bioburden testing on the components used to assemble the syringes.Continuous online monitoring of the wfi (water for injection) water quality (note: this is for toc (total organic carbon) and conductivity, not microbial.Weekly bioburden and endotoxin testing of the usp (united states pharmacopeia) purified water and wfi systems.Weekly endotoxin testing of the pure steam system.Each sterilizer is thoroughly validated before use for posiflush sterilization.A three year review of complaints, environmental monitoring results, water testing results, and quality notifications have confirmed that the serratia marcescens organism has not been recovered in the (b)(4) plant.In addition, the posiflush product produced within (b)(4) is terminally sterilized through steam sterilization which is requalified annually.Furthermore, there is no morphological evidence of serratia marcescens ability to survive the steam conditions used to sterilize the posiflush product manufactured in the (b)(4) plants.For validation, bd (b)(4) & bd (b)(4) uses the spore forming, rod-shaped, gram-positive, thermophile geobacillus stearothermophilus as our biological indicator, along with bacillus atrophaeus.The formation of endospores has not been reported in any species of serratia, thus its resistance to steam sterilization is scientifically remote, further it is a motile organism with growing temperatures ranging from 5¿40 °c, which is well below our >121 °c processing temperatures.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause is undermined.A capa has been opened for further investigation.
 
Event or Problem Description
It was reported that a (b)(6) female with a diagnosis of hodgkin¿s lymphoma and a picc line in place had a positive blood culture for serratia marcescens on (b)(6) 2018.The patient was treated and discharged.The pfge patterns for this incident are indistinguishable from the original co outbreak pattern.The device implicated with this incident is a 5 ml bd posiflush¿ normal saline syringe, in 10 ml syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5 ML BD POSIFLUSH¿ NORMAL SALINE SYRINGE, IN 10 ML SYRINGE
Common Device Name
PREFILLED SALINE FLUSH SYRINGE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
MDR Report Key7594801
Report Number1911916-2018-00292
Device Sequence Number16194227
Product Code NGT
UDI-Device Identifier30382903065456
UDI-Public30382903065456
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K141311
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Catalogue Number306545
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2018
Initial Report FDA Received Date06/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age16 YR
-
-