JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZKB00
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Blurred Vision (2137); Halo (2227); No Code Available (3191)
Date of Event 04/04/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Explant, photopsia, unexpected postop refraction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
A report was received that a patient came in the surgery center on (b)(6) 2018 due to experiencing halos while driving at night and blurred vision in the right eye (od) post implant of a model zkb00 intraocular lens (iol).Per the account, the doctor examined the patient's od and deemed that the iol was subluxed with refractive shift and intolerable photopsia.Consequently, the doctor recommended the iol be exchanged.The lens was explanted on (b)(6) 2018.It was noted that there were no incision enlargement, no vitrectomy, no suture required and there was no injury post-surgery.No additional information provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 6/26/2018.Device returned to manufacturer? yes.The product was received in a small container.Visual inspection shows the lens was damaged at the edge of the optic area.Investigation of the return sample and the condition of the return sample do not suggest the damage was introduced during manufacturing.The manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.As a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7595108
Report Number9614546-2018-00573
Device Sequence Number1277922
Product Code MFK
UDI-Device Identifier05050474560109
UDI-Public(01)05050474560109(17)200314
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/20/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date03/14/2020
Device Model NumberZKB00
Device Catalogue NumberZKB00U0210
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 05/16/2018
Supplement Date Received by Manufacturer09/03/2018
Initial Report FDA Received Date06/13/2018
Supplement Report FDA Received Date09/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Date Report Sent to FDA01/01/2005
-
-