| Model Number |
MMT-723NAL |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed motor error and failed battery test.Customer's blood glucose level was not reported.The customer was unable to rewind the insulin pump due to alarm.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the unexpected restart error test, prime test, excessive no delivery test and occlusion test due to motor error alarm.Device received with normal operating current.Device passed self test.Device passed off no power test.
|
| |
|