| Catalog Number |
LA6AL10 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Material Deformation (2976)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
05/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A 6f launcher guide catheter was attempted to be used in a procedure to treat a moderately tortuous lesion in the rca.There was no damage noted to the device packaging.There were no issues when removing the device from the packaging.The device was inspected with no issues.The device was prepped per ifu with no issues.The lesion was pre-dilated to 22 atm using a non-medtronic sc balloon catheter.Resistance was encountered when advancing the device.It is reported that the physician tried to deliver a non-medtronic drug eluting stent through the guide catheter but the stent got lodged and would not move.The stent tip did not exit the tip of the guide catheter at any stage.The physician then removed the whole system, by withdrawing the guide with the stent inside, resulting in a dissection of the rca.No intervention was required.The procedure was then abandoned, and no additional treatment was performed.The physician is unsure what the adverse event is related to.The patient is reported as being stable post-procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: as received the catheter is engaged to a wire, y adaptor and a drug eluting stent.Part of the wire and drug eluting stent exited the tip of the launcher, approx.4cm for the drug eluting stent and 19cm for the guide wire from the tip, as received.The launcher catheter tip was noted to be folded inwards.No other visual damage as kinks or crushes were noted to the catheter shaft.The catheter was cleaned with running warm water and submerged in the water for 15 minutes to soften the dried blood.The catheter was then separated from the other devices.Upon separation, severe thrombus was noted on the distal stent struts, the thrombus was then dissolved with water resulting in exposure of the proximal stent struts.The stent struts were examined under magnification, damage was detected on the struts and the area adjacent to the struts.Actual physical measurement for outer and inner diameter met specifications.Dye testing confirmed the presence of coating present in the lumen of the catheter.No damage to the catheter inner lumen was noted.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|