| Model Number |
H7493952838400 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product. device evaluated by mfr: promus premier us mr 4.00 x 38mm stent delivery system was returned for analysis.The distal portion of the device was not returned.This includes the wire exchange port, shaft polymer extrusion, balloon, stent and tip.The stent was not returned for analysis as the distal portion of the device was detached and not returned.The manufacturing crimp data for this device was reviewed and there were no anomalies noted.The maximum stent profile was within max crimped stent profile measurement.A visual and tactile examination of the device found multiple hypotube kinks.An examination of the shaft polymer extrusion revealed a midshaft break 1070 mm distal from the distal end of the strain relief exposing the core wire.The portion of the device distal of the break site was not returned.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2018.A 4.00x38mm promus premier¿ drug-eluting stent was received with no issue reported.However, upon review of this device, mid-shaft broken and stent detachment/separation occurred.
|
| |
|
Search Alerts/Recalls
|