Model Number M006192132090
Medical Device Problem Codes Break (1069); Calcified (1077); Difficult to Remove (1528)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.It was reported the device was not used past its expiry.(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a polaris¿ ultra stent was removed during a ureteral stent placement procedure performed in the ureter.The date of the procedure was not reported.It was reported that the stent was implanted around (b)(6) 2018.According to the complainant, during stent removal, which occurred around (b)(6) 2018, the stent was found to be calcified.Strong resistance was encountered when attempts were made to remove the calcified stent and the distal end of the stent detached inside the kidney.There was reportedly a stone like coral attached to the stent.The stone inside the kidney was broken up and removed with the use of laser.The detached stent piece was removed by using a flexible scope.There were no patient complications reported a result of this event.The patient's condition at the conclusion of the procedure was reported to be "stable".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARIS¿ ULTRA
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
MDR Report Key7596064
Report Number3005099803-2018-01927
Device Sequence Number179103
Product Code FAD
UDI-Device Identifier08714729124436
UDI-Public08714729124436
Combination Product (Y/N)N
PMA/510(K) Number
K010002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM006192132090
Device Catalogue Number192-132-09
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2018
Initial Report FDA Received Date06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-