MAUDE Adverse Event Report: BECTON, DICKINSON & CO. BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 368056
Medical Device Problem Codes Delivered as Unsterile Product (1421); Component Missing (2306)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for missing stopper with the incident lot was observed.A review of the manufacturing records was completed for the incident lot and no issues were identified.Investigation conclusion: based on evaluation of the customer photos, the customer¿s indicated failure mode for missing stopper with the incident lot was observed.Root cause description: based on the investigation, the most likely root cause was determined to be related to the manufacturing process where stopper was not aligned properly on the lid when the 300 batch tray was evacuated.The stopper would then fall off when the tube was transferred to the tulip bed.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that bd vacutainer® pst¿ gel and lithium heparinn (lh) 56 units blood collection tubes were missing the stopper in the package.No report of injury or medical intervention.No report of blood exposure to mucous membrane.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key7596874
Report Number1917413-2018-01725
Device Sequence Number19388450
Product Code JKA
UDI-Device Identifier50382903680561
UDI-Public50382903680561
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K945952
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2018
Device Catalogue Number368056
Device Lot Number7214715
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2018
Initial Report FDA Received Date06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-