An event regarding revision due to instability with malposition of the baseplate involving a triathlon insert was reported.The event was confirmed through clinician review of the medical records provided.Method & results: - device evaluation and results: material analysis report (mar) indicated: damage consistent with the explantation process and contact against the femoral component were observed on the insert.Impression markings were observed on the distal surface of the insert, consistent with contact against the baseplate.Delamination burnishing, scratching and third-body indentations being observed on the insert.Delamination, burnishing, scratching and third-body indentations were observed on the insert.These are common damage modes of uhmwpe.Explantation damage and markings consistent with contact against the baseplate were also observed on the insert.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.- medical records received and evaluation: baseplate malposition in excessive posterior tibial slope and excessive proximal tibial bone removal requiring implantation of a very thick 19-mm cr bearing have rendered the pcl dysfunctional with severe instability as outcome requiring revision where the problem was one partly solved by implantation of a thicker 22-mm cs bearing.- device history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.- complaint history review: there have been no other similar events for the lot referenced.Conclusion: damage consistent with the explantation process and contact against the femoral component were observed on the insert.Impression markings were observed on the distal surface of the insert, consistent with contact against the baseplate.Delamination burnishing, scratching and third-body indentations being observed on the insert.Delamination, burnishing, scratching and third-body indentations were observed on the insert.These are common damage modes of uhmwpe.Explantation damage and markings consistent with contact against the baseplate were also observed on the insert.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.The medical review indicated that baseplate malposition in excessive posterior tibial slope and excessive proximal tibial bone removal requiring implantation of a very thick 19-mm cr bearing have rendered the pcl dysfunctional with severe instability as outcome requiring revision where the problem was one partly solved by implantation of a thicker 22-mm cs bearing.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|