| Catalog Number |
6260-5-232 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Injury (2348); Reaction (2414)
|
| Date of Event |
05/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Additional information has been requested.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The patient underwent tha surgery with centpillar on (b)(6) 2009.The patient claimed abnormal sound from the left hip and a dislocation of the metal head was confirmed.Revision surgery was performed on (b)(6) 2018, and an insert, the metal head and the stem were replaced.Wear of the stem neck was noticed.Also, inside of the patient's hip joint was black by metal abrasion powder.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding wear, altr & disassociation involving a metal head was reported.The event of wear and disassociation was confirmed.Altr was not confirmed.Method & results: -device evaluation and results: visual inspection: visual inspection was performed as part of the material analysis report (mar).Damage was also observed on the taper of the head, likely from interacting with the trunnion of the stem.Debris was also observed on the taper of the head.Dimensional & functional inspection: not performed as there is evidence from the visual inspection that the device is damaged.Material analysis: a material analysis has been performed.The report concluded: damage was observed on the stem trunnion and v40 head taper.This damage was consistent with wear mechanisms due to the head taper and stem trunnion losing their taper lock.Debris was observed in the taper of the head.Eds showed the stem and head chemistries were consistent with their drawing callouts and the debris was consistent with a corrosion product, biological material, material transfer from the stem, and an oxide.-medical records received and evaluation: a review of the provided medical records and x-rays by a clinical consultant indicated: review of these records confirms dissociation of the femoral head from the trunnion of the femoral stem with subsequent damage to the trunnion and head occurred.These events were caused by failure the taper lock between the head and trunnion.-device history review: could not be performed as lot code information was not provided.-complaint history review: could not be performed as lot code information was not provided.Conclusion: the clinician confirmed.These events were caused by failure the taper lock between the head and trunnion, however, the root cause could not be determined.
|
| |
|
Event or Problem Description
|
|
The patient underwent tha surgery with centpillar on(b)(6) 2009.The patient claimed abnormal sound from the left hip and a dislocation of the metal head was confirmed.Revision surgery was performed on (b)(6) 2018 ,and an insert, the metal head and the stem were replaced.Wear of the stem neck was noticed.Also, inside of the patient's hip joint was black by metal abrasion powder.
|
| |
|