Model Number MMT-7008B
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem Skin Irritation (2076)
Event Date 05/17/2017
Event Type  malfunction  
Manufacturer Narrative
Inspected 1 opened/used sensor and performed continuity resistance test and sensor failed test.Found cannula bent unable to confirm customer received sensor in said condition due to customer returned opened/used.
 
Event Description
The customer reported via phone call that they had sensor reading discrepancies and the insulin pump went into threshold suspend.The customer stated that the sensor was reading 58, but blood glucose value was 168 mg/dl.The customer was advised about insertion and calibration best practices.The sensor was removed and found the cannula was bent.The sensor was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008B
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire st.
northridge, CA 91325-1219
MDR Report Key7597157
MDR Text Key111317394
Report Number2032227-2018-05427
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-7008B
Device Catalogue NumberMMT-7008B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received05/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age62 YR
Patient Weight177
-
-