| Model Number |
CRPLUS |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer was provided with a replacement device.Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had all three icons (charge pak, attention and service wrench) present on its readiness display.The three icons being illuminated indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and verified the reported issue.Physio determined that the cause of the reported issue was that the hybrid layer capacitor (hlc) batteries located on the analog pcb assembly could no longer hold a charge.The hlc batteries would no longer hold a charge because they were damaged after having previously depleted to zero (0).
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had all three icons (charge pak, attention and service wrench) present on its readiness display.The three icons being illuminated indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use associated with the reported event.
|
| |
|