• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ZENITH AAA ENDOVASCULAR GRAFT AAA ANCILLARY COMPONENT MAIN BODY EXTENSION; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ZENITH AAA ENDOVASCULAR GRAFT AAA ANCILLARY COMPONENT MAIN BODY EXTENSION; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number G32183
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/27/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event or Problem Description
It was reported that while the zenith aaa endovascular graft aaa ancillary component main body extension was being deployed, the device was pressed against the right side of the aortic wall and only deployed properly on the left side.The physician could not determine if the device had been positioned correctly below the lowest renal artery, so they decided to remove the graft, which was re-sheathed and removed successfully.Another backup device was prepared, but the physician decided to make a second attempt at deploying the original graft.With some difficulty, the new graft was deployed into the existing zenith graft, but it was approximately 4-5 mm lower than intended.The physician continued and implanted a limb extension.A final digital subtraction angiogram was performed, showing a successful result with no type i endoleaks.There were no injuries or additional procedures reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZENITH AAA ENDOVASCULAR GRAFT AAA ANCILLARY COMPONENT MAIN BODY EXTENSION
Common Device Name
MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7597271
Report Number1820334-2018-01773
Device Sequence Number11029071
Product Code MIH
UDI-Device Identifier10827002321837
UDI-Public(01)10827002321837(17)151001(10)4543965
Combination Product (Y/N)N
Initial Reporter CountryGB
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG32183
Device Catalogue NumberESBE-30-39
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2018
Initial Report FDA Received Date06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-