• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97714
Device Problems Failure to Interrogate (1332); Improper or Incorrect Procedure or Method (2017); Low Battery (2584); Battery Problem (2885); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received regarding a patient implanted with a neurostimulator (ins) used for spinal pain.Manufacturer representative reported the patient¿s battery was overdischarged due to not charging.It was reported that they attempted to communicate with the battery but were unable to.The patient did not have their recharger and made an appointment (b)(6) 2018 to trickle charge.Additional information was received from a manufacture representative (rep) on (b)(6) 2018.The battery was overdischarged.The rep performed a trickle charge three times but they were unsuccessful.The patient felt better so they stopped recharging which led to the overdischarge.The patient would make an appointment to be seen regarding an implantable neurostimulator (ins) change.It was unknown if surgical intervention was planned.The issue was not resolved but the patient was noted to be alive with no injury.Additional information was received from the health care provider (hcp) via a manufacturer representative on 2018-06-11 confirming that a surgical intervention was planned but had not been scheduled.No further complications were reported.
 
Manufacturer Narrative
A supplemental report will be sent when analysis results are available.H6: contains all applicable codes at this time.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was confirmed that the device was replaced on (b)(6) 2018.The patient recovered without sequela.No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Analysis determined that the battery had a reduced capacity.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7597371
MDR Text Key110941434
Report Number3004209178-2018-13329
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169109483
UDI-Public00643169109483
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/14/2018
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/11/2018
Initial Date FDA Received06/13/2018
Supplement Dates Manufacturer Received07/11/2018
09/10/2018
08/08/2018
Supplement Dates FDA Received07/13/2018
09/18/2018
10/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-