| Catalog Number |
222287 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/31/2013
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was received and reviewed.Visual observations reveal the sutures are wrapped around and cut through the anchor.The distal tip of the anchor looks deformed consistent with excess tension when the suture is being pulled back.The free limbs of the suture should be released after driving the anchor in to release tension.There is evidence that the deformity of the anchor was caused by excess tension from pulling suture back.This type of failure can be attributed to user technique.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no complaints of any type for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device will be disposed through proper channels.Udi: (b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
| |
|
Event or Problem Description
|
|
It was reported an arthrex bio-corkscrew anchor and healix knotless was used as the lateral row anchor.Anchor did not bite to the bone and pulled out of the bone.The anchor did no disengage from the inserter and is stuck to the inserter.The suture had to be cut from the anchor for the surgeon to complete the procedure.
|
| |
|
Search Alerts/Recalls
|