SMITH & NEPHEW, INC. GNS II CON INS SZ 3-4 15MM; PRSTHSS,KNEE,PATELLOFEMOROTIBIAL,SM CNSTRND,CEMENTED,POLYMER/METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
71420528 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
05/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during a poly exchange/washout yesterday, doctor was unable to get both size 3-4 15mm constrained inserts to lock into the revision baseplate.One seemed to be damaged while impacting.Surgery was delayed more than 2 hours.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint device was not returned.A clinical evaluation was conducted and several attempts have been made to obtain clinical/medical documents to no avail.Without supporting clinical/medical documents a thorough investigation cannot be performed.Should information become available this complaint can be re-assessed.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
|
| |
|