• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No parts have been returned to the manufacturer for analysis.A medtronic representative went to the site to test the equipment.Testing revealed that the system functioned as intended.
 
Event or Problem Description
Medtronic received information regarding a navigation system being used during a transforaminal lumbar inter-body fusion (tilf).It was reported that when the site navigated on the imaging system, navigation was 2mm off the skin in the sagittal view.This event occurred intra/perioperatively with less than one hour delay to surgical time.There was no impact on patient outcome.
 
Additional Manufacturer Narrative
Correction: unique device identification (udi), device manufacture date updated to proper value.
 
Additional Manufacturer Narrative
There was no injury associated with the inaccuracy.The issue was resolved by the healthcare provider adjusting the inaccuracy and also adjusting the frame movement from placement close to the skin.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
MDR Report Key7597787
Report Number1723170-2018-02756
Device Sequence Number1133867
Product Code HAW
UDI-Device Identifier00613994724014
UDI-Public00613994724014
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2018
Supplement Date Received by Manufacturer06/26/2018
06/26/2018
Initial Report FDA Received Date06/13/2018
Supplement Report FDA Received Date06/29/2018
02/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
Patient Weight90
-
-