| Model Number |
801413-1 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Fire (1245)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The technician has tested wires to make sure there aren't any other short circuits with the machine.He replaced the capacitor board, tubehead, and cable 58.The machine is working now.(b)(4).
|
| |
|
Event or Problem Description
|
|
The aps technician, (b)(6), was at the office working on the machine and replaced cable 58 on the machine.When the machine was powered on, the capacitor board started crackling and then burst into flames.The technician had to put the fire out with his sweater.The aps technician was not physically injured.
|
| |
|