• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAKO DENMARK A/S HERCEPTEST FOR THE DAKO AUTOSTAINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAKO DENMARK A/S HERCEPTEST FOR THE DAKO AUTOSTAINER Back to Search Results
Catalog Number K5207
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/09/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of dako's internal audit and continuous improvement process, dako performed an internal review of complaints beginning with those initiated january 1, 2015.This report is based on that review as complaint (b)(4) was determined to be mdr reportable.It is therefore being filed with the fda retrospectively.This is an initial and final report combined.The investigation for this case was closed in 2016, but during review the case has been reopened and reassessed as part of the internal review.
 
Event or Problem Description
Device involved in this complaint: · k5207, herceptest for dako autostainer.Summary: inconsistent staining on product k5207/herceptest for the dako autostainer, lot no.20029667, resulted in a false negative.The customer reran the patient tissue on the herceptest assay which confirmed the negative test result.The customer also ran a fish test on the patient tissue which resulted in a positive result.No misdiagnosis occurred as the patient was diagnosed according to the fish testing.Assessment: the customer ran a fish on the specific patient tissue after the ihc tissue test and retest resulted as negative for her2.The fish test indicated the tissue was positive.The customer indicated they thought their procedure was the cause of negative tests.A service engineer was sent on site to work with the customer.The service engineer confirmed their procedure.No root cause or likely root cause was identified.The serial number of the autostainer was not recorded in the complaint.As part of its internal review, dako analyzed whether there were any additional complaints logged on the customer's lot number and no other complaints were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HERCEPTEST FOR THE DAKO AUTOSTAINER
Common Device Name
HERCEPTEST FOR THE DAKO AUTOSTAINER
Manufacturer (Section D)
DAKO DENMARK A/S
produktionsvej 42
glostrup, 2600
DA  2600
MDR Report Key7598042
Report Number9610099-2018-00002
Device Sequence Number11201242
Product Code MVC
UDI-Device Identifier05700572031994
UDI-Public05700572031994
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
P980018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberK5207
Device Lot Number20029667
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/09/2016
Initial Report FDA Received Date06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-