| Model Number |
H965430600 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/22/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.(b)(4).The device has not been received for analysis; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that a wallstent biliary uncovered stent was to be used to treat a malignant stricture in the hilar region of the common bile duct during a stent placement procedure performed on (b)(6) 2018.Reportedly, the patient's anatomy was not dilated prior to stent placement.According to the complainant, during stent deployment, the stainless steel hub suddenly twisted, and causing 80- 90% of the stent to be deployed outside the papilla.The stent was removed from the papilla with forceps and the procedure was completed with another wallstent biliary stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that a wallstent biliary uncovered stent was to be used to treat a malignant stricture in the (b)(6) of the common bile duct during a stent placement procedure performed on (b)(6) 2018.Reportedly, the patient's anatomy was not dilated prior to stent placement.According to the complainant, during stent deployment, the stainless steel hub suddenly twisted, and causing 80- 90% of the stent to be deployed outside the papilla.The stent was removed from the papilla with forceps and the procedure was completed with another wallstent biliary stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
A wallstent biliary delivery system was received for analysis; the stent was not returned.Visual examination of the returned device found that the stainless steel tube broken and kink.The clear exterior tube at the distal end and the inner member were found kinked.No other issues with the device were noted.The noted damages on the returned device were consistent with the excessive force being applied to the delivery system.The investigation concluded that procedural factors as the handling of the device and normal procedural difficulties encountered during the procedure could have possibly affected the device performance and its integrity contributing to the failure experienced by the customer.Therefore, the most probable root cause is operational context.A labeling review was performed and from the information available, this device was used per the directions for use (dfu)/product label.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.A search of the complaint database confirmed that no other similar complaints exist for the specified lot.
|
| |
|