| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Pka end effector handpiece does not work.The magnet was missing from the trigger.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: it was reported that the end effector was non- functional.Product evaluation and results: the product was unavailable for inspection as the product was not returned.Product history review: review of the device history records indicate 20 devices were manufactured and accepted into final stock on 12/17/2015 with no reported discrepancies.Complaint history review: review of complaints within the trackwise database for p/n 111758, l/n 19220315 show three 4 other complaints related to the reported failure.The pr's are (b)(4).Conclusions: the failure could not be determined as the product was not available for inspection.If additional information is received the complaint will be reopened.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there have been an nc and capa associated with the product and failure mode reported in this event.This is nc 1454143 and capa 1454425 h3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
Pka end effector handpiece does not work.The magnet was missing from the trigger.
|
| |
|