• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523NAH
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/28/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Failure analysis: the unit had intermittent bolus express and escape buttons response due to corroded keypad traces.However, unit was received with all operating currents within specifications and passed functional testing including the rewind test, self-test, unexpected restart alarm error test, basic occlusion test, occlusion test, prime/compromised force sensor system test, excessive no delivery test and displacement test.No moisture damage on electronic assembly during visual inspection.Unit had minor scratched lcd window, cracked lcd window, cracked battery tube threads, cracked reservoir tube lip, stained address/serial number label and stained end cap sticker.
 
Event or Problem Description
A customer reported via phone call indicating that their insulin pump was damaged with water after the customer went into the swimming pool with the device.The customer¿s blood glucose level was 111 mg/dl at the time of the incident.The customer was informed that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the back up plan.The customer sent the product back for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7598813
Report Number3004209178-2018-80960
Device Sequence Number10809014
Product Code OYC
UDI-Device Identifier00643169202313
UDI-Public(01)00643169202313
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Patient Family Member or Friend
Type of Report Initial
Report Date (Section B) 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAH
Device Catalogue NumberMMT-523NAH
Device Lot NumberA9523NAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2017
Initial Report FDA Received Date06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age4 YR
Patient Weight20
-
-