Failure analysis: the unit had intermittent bolus express and escape buttons response due to corroded keypad traces.However, unit was received with all operating currents within specifications and passed functional testing including the rewind test, self-test, unexpected restart alarm error test, basic occlusion test, occlusion test, prime/compromised force sensor system test, excessive no delivery test and displacement test.No moisture damage on electronic assembly during visual inspection.Unit had minor scratched lcd window, cracked lcd window, cracked battery tube threads, cracked reservoir tube lip, stained address/serial number label and stained end cap sticker.
|