| Model Number |
407452 |
| Medical Device Problem Codes |
Pacing Problem (1439); Under-Sensing (1661); Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
|
| Health Effect - Clinical Codes |
Syncope (1610); Bruise/Contusion (1754); Non specific EKG/ECG Changes (1817); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/24/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical product: product id: 5076-52 lead, implanted: (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of the right ventricular (rv) lead, the patient had multiple five second pauses.Evaluation revealed an increase in rv lead threshold.An x-ray confirmed that the rv lead had dislodged.The rv lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had passed out and bruised their face and knees.No further patient complications have been reported as a result of this event.
|
| |
|