• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INC. COROENT INTERLOCK SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INC. COROENT INTERLOCK SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL Back to Search Results
Model Number 6790225
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/01/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product has been returned for evaluation as it remains in-situ.No x-rays were provided to confirm the alleged event.As per reporter there is questionable fusion on c3/4 and c5/6 levels.Labeling review: warnings, cautions and precautions ".These devices can break when subjected to the increased load associated with delayed union or nonunion.".Potential adverse events and complications ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s); nonunion or delayed union.".".Patient education: the patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed.".
 
Event or Problem Description
On (b)(6) 2016 patient underwent an anterior cervical discectomy and fusion procedure.Reportedly, six months post-operatively, incidental finding of a fractured screw at the c6 vertebral body.Patient is asymptomatic and no revision procedure will be performed at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COROENT INTERLOCK SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk blvd
san diego CA 92121
MDR Report Key7599518
Report Number2031966-2018-00099
Device Sequence Number1133772
Product Code OVE
UDI-Device Identifier00887517361608
UDI-Public00887517361608
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K102547
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number6790225
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2018
Initial Report FDA Received Date06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-