JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that tecnis toric intraocular lens (model zct150, +25.0 diopter) was removed due to capsular tear/ vitrectomy.There was no patient injury and the lens was not implanted.No further information provided.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7599825
Report Number9614546-2018-00580
Device Sequence Number16660899
Product Code HQL
UDI-Device Identifier05050474552241
UDI-Public(01)05050474552241(17)181110
Combination Product (Y/N)Y
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/07/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/10/2018
Device Model NumberZCT150
Device Catalogue NumberZCT150U250
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 05/17/2018
Supplement Date Received by Manufacturer08/21/2018
Initial Report FDA Received Date06/14/2018
Supplement Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-